Product Details:
|
Usage: | Silicone Rubber Corrosion Resistance Testing Machine | Test Sample: | Electrical Insulating Materials |
---|---|---|---|
Standards: | IEC60587-2007, ASTM D2303 | Number Of Tests: | 5 Groups |
Power: | AC 220V/50HZ | Test Voltage: | 100~6000V Adjustable |
Test Time: | 6 Hours | Warranty: | 12 Months |
Highlight: | AC220V Erosion Tester,ASTMD2303 Tracking Erosion Tester,IEC60587 Erosion Tester |
IEC 60587 Electrical Insulating Materials Resistance To Tracking And Erosion Test Machine
IEC60587 ASTMD2303 Electrical Tracking Mechanism
Technical Description:
The mechanism of electrical impingement on the epoxy resin insulating material using a high voltage of 3.5 kV AC with a frequency of 50 Hz causes the surface of the epoxy resin insulating material to burn.
Five groups of samples are powered on for six hours. When the sample is burned out, the equipment will give an alarm, and the results can be printed after the test.
Model: DX8427
Standard:
IEC60587-2007, ASTMD2303.
Parameters:
Track: partially conducting path created by localized deterioration on the surface of an insulating material;
Erosion, electrical: loss of material by leakage current or electrical discharge;
Name | DX8427 High Voltage Tracking Index Tester |
Box |
≥ 0.5 cubic meters |
Test specimens dimensions | 50mm*120mm, thickness 6mm |
Drop size |
44 to 50 drops / cm3 |
Power supply |
AC 220V, 50Hz |
Electrical load |
100 ~ 6000V adjustable |
Drip height |
30mm ~ 40mm adjustable |
Electrode Pressure |
1.00N±0.05N |
Flow range | 0.00185 to 2 ml / min |
Working speed | 0.1 to 15 rev / min |
1. DC voltage: display the voltage of DC test, which can be realized by adjusting voltage regulator;
2. AC voltage: display the voltage of AC test, which can be realized by adjusting voltage regulator;
3. DC current (1 ~ 5): there are five groups of DC current display, display the current value of five groups of DC test, the set value is 60mA;
4. AC current (1 ~ 5): there are five groups of AC current display, display the current value of five groups of AC test, the set value is 60mA;
5. Reset: 1. When the tester alarms, press this button to stop the alarm; 2. Press this button to stop the test, and all the current values in the main test screen are cleared;
6. Power supply: it can control the on-off of power supply and has the function of leakage protection;
7. Resistance gear: select the gear of series resistance value;
8. Preferred test voltages are 2.5KV, 3.5KV and 4.5KV;
Contact Person: Sophia Su
Tel: +86-13266221899